How do I calibrate my Urisys 1100?
If the Urisys 1100® analyzer is in Ready-to- Measure mode, press the left function key to select “Calibr.” Next, the message “START CALIBRATION” is displayed. 2 Place the calibration strip, with pads facing upward, on the test strip tray. About 2 mm of strip must be held under the clip. 3 Press the START button.
What is Urisys?
The Urisys 1100 urine analyser is a semi-automated hand held urine testing analyser for a workload of up to 50 urine samples per day and is designed to improve workflow efficiency in smaller labs, doctor offices and decentralized lab settings like point-of-care.
What is urine analyzer?
A urine analyser is a device used in the clinical setting to perform automatic urine testing. The units can detect and quantify a number of analytes including bilirubin, protein, glucose and red blood cells.
What is the principle of Urisys 1100 urine analyzer?
The Urisys 1100® Urine Analyzer standardizes urine test strip results by eliminating factors known to affect visual evalu- ation of urine test strips, such as; variable lighting conditions at the workplace, individual skill levels at matching test strip pad colors, failure to keep the prescribed testing times, and …
How do you read hCG Clinitest?
The CLINITEST hCG test
- When the urine hCG level is ≥ 25 mIU/mL, it indicates a positive result.
- When the urine hCG level is ≤ 5 mIU/mL, it indicates a negative result.
- When the urine hCG level is between 5 and 25 mIU/mL, it indicates a borderline result.
How do you read a urine report?
Normal values are as follows:
- Color – Yellow (light/pale to dark/deep amber)
- Clarity/turbidity – Clear or cloudy.
- pH – 4.5-8.
- Specific gravity – 1.005-1.025.
- Glucose – ≤130 mg/d.
- Ketones – None.
- Nitrites – Negative.
- Leukocyte esterase – Negative.
What are 3 things that a urinalysis tests for?
Overview. A urinalysis is a test of your urine. It’s used to detect and manage a wide range of disorders, such as urinary tract infections, kidney disease and diabetes. A urinalysis involves checking the appearance, concentration and content of urine.
What is the testing procedure for the urisys 1100 urine test?
Testing Procedure 1. User assembles testing supplies q Urisys 1100®Urine Analyzer q Chemstrip Urine Test Strips q Chemstrip Urine Analyzer Calibration Strips q Patient specimen q Disposable latex gloves and laboratory coat q Absorbent paper 2.
Is urisys 1100 a Roche product?
COBAS, Urisys 1100®and CHEMSTRIP are trademarks of Roche. 41 Skills Checklist Urisys 1100™ Urine Analyzer The trainer should check each activity as it is demonstrated. All activities must be completed before the trainee is certi- fied to use the Urisys 1100®Urine Analyzer.
How do I clean and disinfect the urisys 1100®urine analyzer?
• Clean and disinfect the exterior of the Urisys 1100®Urine Analyzer by using a cloth dampened with a suitable disin- fectant such as 10% bleach or isopropyl alcohol. Be sure that no liquid enters the analyzer.
How do I know when my urisys 1100®analzer is ready to start reading?
The Urisys 1100®Analzer is ready to begin reading test strips when the message “INSERT STRIP” is displayed. Note for those testing with Chemstrip 5 OB or Chemstrip 7 Test Strips: If you have inserted the correct test strip tray (Type N), the message “E9 WRONG TRAY!” will be displayed after the self check.